Though no medical/surgical intervention was required to preclude a serious injury in this event, there have been previously reported events involving a similar device that resulted in the need for medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.Therefore, this event meets criteria for reportability per 21 cfr part 803.As received, the handpiece could not support the reported complaint and a root cause could not be determined.There was no evidence of the cap unscrewing under testing conditions where the cap was affixed with the specified torque wrench.The cap only moved when it was unscrewed by half a revolution during testing.Multiple dimensions on the head and cap were checked by production personnel and all were found to be within specification.The handpiece was then microscopically evaluated by quality personnel.Microscopic evaluation revealed significant debris within the cap and head cavities and set components.Some rubbing marks were also seen within the head cavity.All components looked dry with no evidence of lubricant.Dentsply is unable to determine how the cap might have come unscrewed as all handpieces are 100% performance tested prior to shipping.
|