MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 3058 |
Device Problems
Failure to Deliver Energy (1211); High impedance (1291); Unable to Obtain Readings (1516)
|
Patient Problems
Therapeutic Effects, Unexpected (2099); Therapeutic Response, Decreased (2271); Complaint, Ill-Defined (2331)
|
Event Type
malfunction
|
Event Description
|
It was reported that patient experienced loss of therapeutic effect.The patient claimed that they did not see good symptom control from date of implant.The patient complained of not feeling stimulation for about a year since implant.The patient had a worsening of urinary symptoms recently.The patient was programmed with 3+ 0- at a voltage of 2.3v.The following impedances were reviewed: c0: 1226 c1:??? c2: 1111 c3: 628, 01 1482 02: >4000 03: 2213.The health care provider wanted to help programming around an elevated impedance value.The patient was programmed to c+ 3- pulse width: 210us rate: 14 hz and confirmed that patient felt stimulation at a voltage of 1v.No outcome was reported regarding this event.A further follow-up is being conducted to obtain this information.A follow-up report will be sent if additional information becomes available.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id 3037, serial# (b)(4), product type: programmer, patient; product id 3 093-33, lot# v508248, implanted: (b)(6) 2010, product type: lead.(b)(4).
|
|