MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS EMAX 2 PLUS MOTOR; MOTOR, DRILL, ELECTRIC - HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS EMAX 2 PLUS MOTOR; MOTOR, DRILL, ELECTRIC - HANDPIECE Back to Search Results
Catalog Number EMAX2PLUS
Device Problems Tear, Rip or Hole in Device Packaging (2385); Material Protrusion/Extrusion (2979); Torn Material (3024)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2014
Event Type  malfunction  
Event Description
It was reported that during a routine maintenance check, it was observed that there was a tear in the cable covering on the motor device.It was further reported that there were no wires exposed.During in-house engineering evaluation it was observed that the cord had been torn at the connector, which exposed the spring and wires.The event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the cord was torn at the connector, which exposed the spring and wires.Therefore, the reported condition was confirmed.It was determined that this was caused by pulling on the cord instead of the connector body to unplug the device from the console.The assignable root cause was determined to be due to misuse, abuse and/or user error.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4520809
MDR Text Key5533449
Report Number1045834-2015-10357
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2014
Initial Date Manufacturer Received 01/13/2015
Initial Date FDA Received02/17/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-