ADVANCED STERILIZATION PRODUCTS STERRAD 100NX CASSETTE; STERRAD 100NX STERILIZER Back to Search Results
Catalog Number 10144
Device Problem Chemical Problem (2893)
Patient Problem Skin Irritation (2076)
Event Date 02/10/2015
Event Type  malfunction  
Event Description
A facility reported a healthcare worker (hcw) came in contact with hyrdrogen peroxide (h202) while removing a used cassette from a sterrad 100nx.The hcw experienced staining on his finger only with no irritation.Hcw was advised to flush the affected area with water.The hcw did not seek any medical attention or treatment.The event isi being reported as a malfunction report subsequent to a serious injury event dated (b)(6) 2014.
 
Manufacturer Narrative
Ni.
 
Manufacturer Narrative
(b)(4).Asp investigation summary: the investigation included a review of the device batch record, supplier product evaluation, trending of the product malfunction code, and system risk analysis (sra).The batch record was not reviewed as the lot number of the cassette was not available.Supplier product evaluation was not performed as the cause of the reported issue is user error.The trend for the product malfunction code of skin reaction was assessed from march 2014 through february 2014.There was no significant trend.The sra indicates the risk associated with improper handling of a cassette is "low." the instructions for use (ifu) of the sterrad® 100nx cassette state: "caution: wear personal protective equipment if handling a used cassette, or any of the cassette case components that may have been subject to liquid leak.This includes a cassette that has been ejected (for any reason) after insertion." the issue has been attributed to user error as the healthcare worker (hcw) handled a used cassette without utilizing proper personal protective equipment (ppe).A customer letter was sent advising the customer to wear proper ppe to avoid this issue in the future.Review of tracking and trending data did not reveal a trend.As a result, root cause was not performed; therefore, no further investigation is necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD 100NX CASSETTE
Type of Device
STERRAD 100NX STERILIZER
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
joaquin kurz
irvine, CA 92618
MDR Report Key4528973
MDR Text Key5507124
Report Number2084725-2015-00040
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071345
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10144
Device Lot NumberN/A
Other Device ID Number10144
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/10/2015
Initial Date FDA Received02/19/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-