Model Number 801764
Device Problems Device Alarm System (1012); Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/20/2015
Event Type  malfunction  
Event Description
It was reported that during use of the device for a cardiopulmonary bypass procedure, the perfusion system lost communications between the pods and the central control monitor (ccm).An audible alarm occurred.After three minutes, the communication came back.The device was not changed out, they were able to use pumps during lost communication time.They did not change the ccm during bypass.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the pt.
 
Manufacturer Narrative
Per the subsidiary site, the customer has made repairs to the perfusion system.The power manager and network interface card (nic) cables were replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
APS 1 (HEART LUNG CONSOLE)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEM CORP
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4535467
MDR Text Key5354537
Report Number1828100-2015-00122
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801764
Device Catalogue Number801764
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/23/2015
Initial Date FDA Received02/12/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-