• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIDMARK CORP JB-70; UNIT, X-RAY, EXTRAORAL WITH TIMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIDMARK CORP JB-70; UNIT, X-RAY, EXTRAORAL WITH TIMER Back to Search Results
Model Number JB-70
Device Problem Installation-Related Problem (2965)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2015
Event Type  malfunction  
Event Description
It was reported to (b)(4) that on (b)(6) 2015, an x-ray unit (jb-70 intraoral dental x-ray system, serial # (b)(4)) separated from its wall mount in a user facility.No injuries were reported.Because the unit was re-installed before the (b)(4) representative visited the user facility, the mounting site was not able to be inspected.However, photographs show that there is continuous wallpaper at the site of mounting, which indicates that at the time of installation, the wall had not been opened to visualize the stud and that the "installation guide" tool had not been used.Because the installation instructions require that the stud be located visually through opening the wall using the installation guide, root cause of the mounting failure is determined to be improper installation, based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JB-70
Type of Device
UNIT, X-RAY, EXTRAORAL WITH TIMER
Manufacturer (Section D)
MIDMARK CORP
1407 barclay blvd.
675 heathrow dr
lincolnshire IL 60069
Manufacturer Contact
nataliya khutoma
675 heathrow dr
lincolnshire, IL 60069
8474159760
MDR Report Key4551490
MDR Text Key5524732
Report Number1423380-2015-00011
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020070
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberJB-70
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/29/2015
Initial Date FDA Received02/25/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/14/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-