Model Number 8070
Device Problem Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
Customer reported multiple tears at the foot end of the canopy.Customer did not provide a date when the issue was discovered.No patient incident or injury was reported.
 
Manufacturer Narrative
Results: evaluation of the returned product found one element missing from the molding zipper of the left window and 8 inches of the material unraveling on the leg of the foot panel side.One element missing would allow the zipper to initially close, but can separate at the location of the missing element if sufficient pressure is applied.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED
Manufacturer (Section D)
J. T. POSEY CO.
arcadia CA
Manufacturer (Section G)
POSEY, S. DE R.L. DE C.V.
ave. ferrocarril no. 16901. bodega 64
colonia ri o tijuana, 3ra. etapa
tijuana 2266 4
MX   22664
Manufacturer Contact
roxana koussa
5635 peck rd.
arcadia, CA 91006
6264433143
MDR Report Key4552834
MDR Text Key5477050
Report Number2020362-2015-00032
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103817
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8070
Device Catalogue Number8070
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/12/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/16/2015
Initial Date FDA Received02/25/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-