Model Number MCP0.0703309 |
Device Problems
Device Displays Incorrect Message (2591); Insufficient Information (3190)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/19/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that one of the rpm 20's ran at full speed momentarily.This happened just after the case while emptying the vent lines.No reported patient effect.(b)(4).
|
|
Manufacturer Narrative
|
A supplemental medwatch will be submitted when additional information becomes available.(b)(4).Both products are components of the hl20 integrated perfusion system cleared under 51(k) k943803.
|
|
Manufacturer Narrative
|
Field safety technician has been investigated and found failure reported.He added the grounding and no further instances since then.Thus the failure could be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
|
|