Though no medical/surgical intervention was required to preclude a serious injury in this event, there have been previously reported events involving a similar device that resulted in the need for medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.Therefore, this event meets the criteria for reportability per 21 cfr part 803.As received, the handpiece could not support the reported complaint and a root cause could not be determined.Cut testing with the handpiece could not be performed as the handpiece stalled immediately when bur was introduced into cut test material.The handpiece was then microscopically evaluated.Minor to moderate corrosion was noted inside cap and head cavities as well as on the set components.Some damage to the rotor blades was also observed.The threads and depth of both the cap and the head of the returned handpiece were also measured.All dimensions were found to be within specification.All set components appeared to lack lubrication.
|