ETHICON INC. GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY; DEVICE, THERMAL ABLATION, ENDOMETRIAL Back to Search Results
Catalog Number TC003
Device Problem Balloon leak(s) (1052)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/10/2015
Event Type  malfunction  
Event Description
It was reported that a patient underwent an endometrial thermal ablation procedure on (b)(6) 2015.During therapy, an error code was received.The surgeon removed the balloon from the patient and it was noticed the balloon was leaking.The case was stopped at that point.There were no adverse patient consequences reported.Additional information has been requested.
 
Manufacturer Narrative
(b)(4).Conclusion: the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Manufacturer Narrative
(b)(4): actual sample was returned for evaluation.Visual and functional inspections were performed.There were no visual defects found on close inspection and no pinholes or other damages were found on the balloon.Functional inspection for pressure was performed and the pressure in the catheter worked as intended on testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY
Type of Device
DEVICE, THERMAL ABLATION, ENDOMETRIAL
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125, co
l salvacar
ciudad juarez 3260 4
MX   32604
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4569678
MDR Text Key5551837
Report Number2210968-2015-02480
Device Sequence Number1
Product Code MNB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue NumberTC003
Device Lot NumberHEMG07
Other Device ID Number10705031050730
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/05/2015
Initial Date FDA Received03/04/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-