Catalog Number 6474526
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/19/2015
Event Type  malfunction  
Manufacturer Narrative
Additional information: functional evaluation confirms pinion and gear interface slippage.Visual and optical examination of the -03 mill gear and -05 mill pinion interfacing features identified significant material plastic deformation.The location and nature of the wear is consistent with instrument usage; this instrument is a single-use instrument.Conclusion: the above observations are consistent with anticipated wear.
 
Event Description
It was reported that the patient underwent a cervical disc arthroplasty procedure.During the surgery while milling, it was found that the cephalad level end plate did not mill properly.The mill was pulled out to be tested and it was found that it was no longer spinning.Another mill had to be used to complete the surgery.No patient complications were reported.
 
Manufacturer Narrative
(b)(4): the device was not returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRYAN CERVICAL DISC SYSTEM
Type of Device
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer Contact
huzefa mamoola
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key4578366
MDR Text Key19701488
Report Number1030489-2015-00469
Device Sequence Number1
Product Code MJO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/17/2022
Device Catalogue Number6474526
Device Lot NumberEM13H038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2015
Initial Date FDA Received03/06/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00038 YR
-
-