After the operating room nurse assembled the handpiece, it was noted by the distributor that instead of taking 1 minute, the handpiece cooling process was taking longer than usual, approximately 3 minutes.After handpiece cooled, the console made a strange sound and the cooling process restarted again automatically on its own, without any user intervention.On this occasion, the handpiece cooling process took only 1 minute and the amplitude and prime tests were completed satisfactorily.Ten minutes into the procedure, the console lights suddenly turned off and the equipment began recalibrating the handpiece, cooling process and auto-tests.The surgery continued for another 5 minutes until the same event happened again.The distributor performed all the troubleshooting measures, but since the situation persisted, the distributor decided to exchange the handpiece for a different one.Cooling and auto-tests were successful.At some point, when the client pressed the pedal in order to fragment the tissue, the amplitude lights on the console board started flickering and the console simply could not even reach 50% force.The surgeon decided not to use the cusa.Additional information was requested and on 02/12/2015, the following was received from the distributor: patient was a (b)(6) male.When the console lights suddenly turned off and the equipment began recalibrating the handpiece after ten minutes into the procedure, it was in run mode.The handpiece was being used on the patient at the time the event happened again.There was no patient injury reported.There was approximately 1 hour surgical delay.There was no patient adverse consequence as a result of the surgical delay, but anesthesia time was prolonged.There was no product replacement.The surgeon continued the surgery with bipolar, dissector, and suction.Surgery was completed.Patient outcome were reported as "have the expected evolution after the removal of tumor".
|