• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGFA HEALTHCARE CORP IMPAX CV OUTBOUND (RESULTS MANAGEMENT 2.08); PICTURE ARCHIVING AND COMMUNICATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGFA HEALTHCARE CORP IMPAX CV OUTBOUND (RESULTS MANAGEMENT 2.08); PICTURE ARCHIVING AND COMMUNICATION Back to Search Results
Model Number IMPAX CV 2.08
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/13/2014
Event Type  malfunction  
Event Description
Agfa submitted mdr report #1225058-2010-00001 to the fda on june 7, 2010 for a site in the us.A 14th occurrence is being reported for the same issue/same device: impax cv results management administration tool (rmat).This is an internal discovery determined during the implementation of the associated problem correction plan, rmat verification, as reported in fda z-2112-10.Within this occurrence are 1737 different study dates in which an individual mdr report will be submitted for each associated study date and medical record number/s (mrn) identified.
 
Manufacturer Narrative
An agfa clinical analyst performed a retro-analysis and reported the findings to agfa service and agfa product quality manager.Agfa's investigation into this occurrence of rmat customizations has revealed that this specific change had the potential to introduce clinical inaccuracies in patient reports.Specifically: original sentence of: "there is fusion of the" has been changed to: "there is aortic valve sclerosis without significant aortic valve stenosis".In this issue, the original sentence intent that there was fusion of the - the assumption is that the fusion would be of the av leaflets and that this would be included in notes or somewhere else in the report.However, the changed sentence is clinically different and does not indicate fusion whatsoever, but instead indicates the presence of av sclerosis without significant stenosis.This is not the intent of the reading physician and could lead to misdiagnosis and/or inappropriate treatment.Potential impact: if the change was made after reports had been generated with the original sentence selected, those reports prior to the change could display incorrect clinical data if viewed in rm or if used as a template for a repeat study on that pt.There has been no reported pt harm for this occurrence.A reportable correction is underway for this issue and has been reported to the fda via reference #z-2112-10.Agfa will follow the rmat post market verification work instructions to correct the sentence finding.Any further investigation for the site described in this report will be documented in the ongoing cfr part 806 reporting.(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPAX CV OUTBOUND (RESULTS MANAGEMENT 2.08)
Type of Device
PICTURE ARCHIVING AND COMMUNICATION
Manufacturer (Section D)
AGFA HEALTHCARE CORP
1 crosswind rd.
misquamicut RI 02891
Manufacturer Contact
deborah huff
10 south academy street
greenville, SC 29601
8644211754
MDR Report Key4611911
MDR Text Key5656775
Report Number1225058-2014-06629
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050228
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 12/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberIMPAX CV 2.08
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/13/2014
Initial Date FDA Received12/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2002
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-