Model Number 105-5096-000 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/21/2015 |
Event Type
malfunction
|
Event Description
|
Medtronic (covidien) received information regarding an incident with one of its devices.During a procedure, the wire (agility 10-wire) that was inside the apollo micro catheter would not be advanced forward or moved backward.The catheter was removed to be inspected and the detachment part was noted to be separated and severely dented.No patient injury was reported as a result of the procedure.
|
|
Manufacturer Narrative
|
The device involved in the event has not been returned for evaluation.Requests have been made for reason for non-return, but without response.The lot history record of the reported lot number showed no discrepancies that might have contributed to the reported experience.(b)(4).
|
|
Manufacturer Narrative
|
The apollo catheter was returned for evaluation.The catheters useable length was measured to be within specification.The catheter was found to be bent and accordion like.The catheter was also found to be elongated with the guidewire protruding from the catheter tip for approximately 34.0cm.The detachable tip was found to be detached and found on the guidewire at approximately 2.8cm from the distal end of the guidewire.The detachable tip was measured to be within specification.The ldpe coupling tube overlap was measured to be within specification.The guidewire was removed from within the catheter lumen with slight resistance.No other anomalies were observed.The lot history records have been reviewed and no quality issues were noted.The tensile value for "low bond - 33.0" was reviewed and was within control limits of historical spc data.Based on the above findings the customer's report was confirmed.However, the cause for the damages could not be determined.All catheters are 100% inspected for damages and irregularities during manufacture.(b)(4).
|
|
Search Alerts/Recalls
|