It was reported that the vns patient¿s device was unable to be interrogated during an office visit on (b)(6) 2015.Several minutes would pass when initiating the interrogation process with the programming system until a failure message would appear.There were no issues found with the wand battery, the connection between the wand and serial adaptor, and the serial cables.The physician re-attempted to interrogate the patient¿s device by rotating the programming wand 45 degrees and ensuring the programming wand was at least 2-3 feet from all other electrical equipment; however, the issues continued.It was noted that data transfer indicator on the programming wand was not blinking when attempting the interrogations.The physician stated that these attempted interrogations were not taking place in the normal office, but no sources of emi were noted in the current surroundings.The patient¿s device was not programmed to high settings and the patient did not have any recent surgeries where electrocautery was used.No further information relevant to the event has been received to date.The issues associated with the programming wand will be reported in the next quarterly alternative summary report.
|