FISHER & PAYKEL HEALTHCARE LTD BI-LEVEL/CPAP INSPIRATORY HEATED CIRCUIT; BTT Back to Search Results
Model Number RT319
Device Problems Split (2537); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/25/2015
Event Type  malfunction  
Event Description
A hospital in (b)(6) reported an rt319 adult breathing circuit was 'split' and caused the humidifier to alarm.They further reported that they had opened the circuit box with a box cutter, but did not think that this had caused the observed damage.No patient consequence was reported.
 
Manufacturer Narrative
(b)(4).The rt319 adult breathing circuit is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Method: the returned rt319 breathing circuit was visually inspected.Results: a cut was observed in the inspiratory limb.Conclusion: the position and direction of the cut indicated that the cut was caused by opening the box with a sharp object, such as a knife or boxcutter.As the customer had informed us that a boxcutter was used to open the circuit box, we can conclude that the damage was caused at this time.All rt319 breathing circuits are visually inspected and pressure tested for leaks before releasing for distribution.Any breathing circuit which fails any of these tests is discarded.The user instructions supplied with the rt319 breathing circuit state: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BI-LEVEL/CPAP INSPIRATORY HEATED CIRCUIT
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9194534000
MDR Report Key4633755
MDR Text Key5613728
Report Number9611451-2015-00157
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT319
Device Catalogue NumberRT319
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2015
Initial Date Manufacturer Received 02/27/2015
Initial Date FDA Received03/25/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-