A medivators field service engineer (fse) observed that the facility's dsd-201 automated endoscope reprocessors (aer) were running the incorrect program while using rapicide opa/28 high-level disinfectant.The program setting had the incorrect temperature for the opa disinfectant.The fse showed the facility the instructions for use for rapicide opa/28, which states the temperature it should be used at.Not using the disinfectant at the correct temperature could result in endoscopes not being properly disinfected, which could lead to cross-contamination.
|
A medivators field service engineer (fse) observed that the facility's dsd-201 automated endoscope reprocessors (aer) were running the incorrect program while using rapicide opa/28 high-level disinfectant.The program was set to the incorrect temperature for the opa disinfectant.The opa temperature was 17c.The fse showed the facility the instructions for use for rapicide opa/28, which states the temperature it should be used at as 25c.Not using the disinfectant at the correct temperature could result in endoscopes not being properly disinfected, which could lead to cross-contamination.The facility checked several scopes and all were contaminated.There is limited information to the total number of patient procedures performed with improperly disinfected scopes.Medivators has not been notified of any patient illness or injury as a result of this incident.This complaint will continue to be monitored in the medivators complaint handling system.
|