• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 5.5MM RESECTOR CUTTER, FORMULA (5BX); BUR, SURGICAL, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 5.5MM RESECTOR CUTTER, FORMULA (5BX); BUR, SURGICAL, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0375562000
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/11/2015
Event Type  malfunction  
Event Description
It was reported that the blade had heated up.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Manufacturer Narrative
The product was returned and the failure mode was confirmed.The cutter was visually inspected for damages, and the shaft at the proximal end has broken off.The proximal plastic hub appears to be melted.Also, the cutting window has evidence of interference.The inner shaft is stuck and cannot be removed.Unable to perform a functional inspection due to the condition of the cutter.The probable root causes for the broken shaft could be lateral pressure was used causing the proximal end to break, or the tip of the burr was stuck causing the shaver to break the proximal end.The probable root causes for the melted hub could be friction due to excessive force.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that the blade had heated up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5MM RESECTOR CUTTER, FORMULA (5BX)
Type of Device
BUR, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
thomas shafer
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key4648921
MDR Text Key5597741
Report Number0002936485-2015-00228
Device Sequence Number1
Product Code GFF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0375562000
Device Lot Number15012CE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/11/2015
Initial Date FDA Received04/01/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/26/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-