It was reported that the patient was seen that day and high lead impedance >10,000 ohms was seen.She had her battery replaced on (b)(6) 2014.X-rays had already been taken but there was nothing clearly discernible from them.These will not be sent in for review.The physician is aware of the recommendation to disable the device but decided to leave the device on.The patient¿s mother did mention that the patient had been manipulating the device after her replacement surgery in (b)(6) 2014.The physician is going to schedule the patient for surgery to fix her vns.He will try re-inserting the pin first.If that doesn't resolve the issue he will revise the patient's lead.On (b)(6) 2015 the patient underwent a full revision.After the new generator was inserted, a systems diagnostic test was run twice with high lead impedance (>10,000 ohms) observed both times, so the lead was also replaced.The physician believes that the original lead coils were not on the nerve.The explanted devices are not retrievable for return.
|