• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA, INC. PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA, INC. PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE Back to Search Results
Model Number 57063/9207010
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/26/2015
Event Type  malfunction  
Event Description
The customer reported to medela customer service that her pump in style transformer housing was cracked open, exposing inner electrical circuitry, which is a safety risk.
 
Manufacturer Narrative
A replacement power adapter was sent to the customer.And requested the defective power adapter be returned for evaluation.The defective adapter has not been received at medela as of 03/27/2015.The product involved in the complaint was not returned for evaluation/investigation at this time.Therefore, no conclusions can be made as to the cause of the event.Customer follow up was not successful.Should additional information or the original product be received, resulting in new, changed, or corrected information, a follow up report will be filed at that time.The issue with a damaged transformer housing for the pump in style device was addressed in investigation (b)(4).The investigation found that the transformers are being damaged during shipment from the manufacturer to medela.This damage is causing the plastic housing to fail prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the transformers to medela is not robust enough to handle all the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process has been modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength has also been increased to further protect the transformers during shipping.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE
Manufacturer (Section D)
MEDELA, INC.
mchenry IL 60050 700
Manufacturer Contact
don alexander
1101 corporate dr.
mchenry, IL 60050-7005
8004358316
MDR Report Key4657487
MDR Text Key15969183
Report Number1419937-2015-00101
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 03/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number57063/9207010
Device Catalogue Number57063/9207010
Device Lot Number3512 REV M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/26/2015
Initial Date FDA Received04/02/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-