• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - CORK ROTALINK? PLUS; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - CORK ROTALINK? PLUS; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H749236310030
Device Problems Occlusion Within Device (1423); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/05/2015
Event Type  malfunction  
Event Description
Same case as mdr id: 2134265-2015-01704.It was reported that the wire broke and was stuck at the tip of the burr.The target lesion was located in a moderately tortuous and severely calcified coronary artery.A 1.50mm rotalink¿ plus and a 325cm rotawire were selected for treatment.Platform testing was performed with the burr outside of the patient.It was then observed that the wire broke as the loop was in a sharp angle.The physician attempted to insert a new wire on the burr however, it was noted that the remainder of the wire was stuck at the distal tip.The procedure was completed with another of the same device.There were no complications reported and the patient's status was good.
 
Manufacturer Narrative
Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK? PLUS
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
Manufacturer (Section G)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4663272
MDR Text Key5683374
Report Number2134265-2015-01703
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Model NumberH749236310030
Device Catalogue Number23631-003
Device Lot Number0017056975
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/09/2015
Initial Date FDA Received04/07/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-