Catalog Number 0250070520
Device Problem Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2015
Event Type  malfunction  
Event Description
It was reported that there was a hair inside the sterile pack.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Manufacturer Narrative
The foreign material failure mode was confirmed on the unit returned.A small strand of hair was found inside the blister package.Most probable root causes are: manufacturing/assembly error, incorrect or inadequate packaging.As per the evidence at hand root cause is a manufacturing error.This was deduced due to current risk documents and investigation.As per current risk documents, the event is within predicted acceptable occurrence and severity thresholds.The risk is within the green low acceptable level.There are mitigating procedures in place.No further actions are deemed necessary at this time the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that there was a hair inside the sterile pack.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRYKEFLOW2 WITH DISPOSABLE TIP (6BX)
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key4664652
MDR Text Key5608332
Report Number0002936485-2015-00278
Device Sequence Number1
Product Code LJH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042457
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250070520
Device Lot Number14288FG2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/12/2015
Initial Date FDA Received04/07/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/03/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-