Catalog Number 0385950008
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2015
Event Type  malfunction  
Event Description
It was found that the inner shaft was hot enough to cause deformation of the proximal plastic hub.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Manufacturer Narrative
The reported device was not received for investigation; therefore the reported failure cannot be confirmed.The complaint will be closed without a detailed investigation report and based on probable root cause.In the event that the device is received, the complaint will be reopened, a full evaluation will be conducted, and the investigation will be updated with the new results.Probable root cause for the reported failure involving this device could have been caused by: manufacturing defect (grease application), high friction due to components interaction, surface not smooth, poor/lack of suction, clogged shaver blade preventing fluid suction and cooling.The product was not returned for investigation therefore the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was found that the inner shaft was hot enough to cause deformation of the proximal plastic hub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5MM ROUND BUR, LONG HIP BUR, FORMULA, 8 FLUTE (5BX)
Type of Device
BUR, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
thomas shafer
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key4666200
MDR Text Key5730461
Report Number0002936485-2015-00279
Device Sequence Number1
Product Code GFF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0385950008
Device Lot Number15050CE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/13/2015
Initial Date FDA Received04/07/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/19/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-