The user reported that the alarm "sample flow rate error on gas module" appeared during clinical use.The issue was intermittent.The attending clinician restarted the multigas module and the alarm disappeared.The device continued to work as specified and expected until the end of surgery.The "sample flow rate error on gas module" is activated when the multigas module measures a sample flow that is too high or too low and will typically be due to an occlusion in the sample tube or the water trap.The patient was not harmed.
|