DENTSPLY PROFESSIONAL MIDWEST TRADITION L NON FIBER OPTIC; AIR-POWERED DENTAL HANDPIECE Back to Search Results
Catalog Number 780044
Device Problems Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
In this event it was reported that cap unscrewed from a tradition handpiece outside of a patient's mouth.The reported complaint did not result in an injury or need for intervention.
 
Manufacturer Narrative
Though no medical/surgical intervention was required to preclude a serious injury in this event, there have been previously reported events involving a similar device that resulted in the need for medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.Therefore, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Manufacturer Narrative
Evaluation determined the root cause was bur end bearing debris.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST TRADITION L NON FIBER OPTIC
Type of Device
AIR-POWERED DENTAL HANDPIECE
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
des plaines IL
Manufacturer Contact
helen lewis
221 w. philadelphia st., ste. 60
susquehanna commerce center w.
york, PA 17401
7178457511
MDR Report Key4669546
MDR Text Key5605078
Report Number1419322-2015-00022
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K963050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 03/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number780044
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/05/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/05/2015
Initial Date FDA Received04/02/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown