• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE OEC MEDICAL SYSTEMS 9900; FLUOROSCOPIC X-RAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE OEC MEDICAL SYSTEMS 9900; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 9900
Device Problems Difficult or Delayed Activation (2577); Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2015
Event Type  malfunction  
Event Description
The customer reported that the system failed to power to a usable state and exhibited a non-recoverable loss of functionality.No patient serious injury or death was reported related to this event.
 
Manufacturer Narrative
A ge service representative performed an onsite investigation.The fse evaluated and replaced the intelligent shutdown pcb and surge suppressor pcb.The fse also found the connector was not seated fully on cable which carries ac power line sensing signal between surge suppressor and systems interface pcbs.The connector was reseated during the service event.The system was tested and found to be working as intended and returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9900
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key4680399
MDR Text Key17575697
Report Number1720753-2015-01476
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2015
Initial Date FDA Received04/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-