(b)(4).Should the information be provided later, a supplemental medwatch will be sent.Additional information was requested and the following was obtained: how did device not work, did clips not deploy or release? dr.Was applying clip on anastomosis site (edge) he started complaining when the instrument (he thought) went 'wobbly' or bending.Where was the bend on the device (was shaft bent)? he could not release clips when he noticed the issue.How was case completed? we had to open another one (ligamax) to proceed with the operation to achieve hemostasis.The analysis results for the el5ml device found that it was returned with the shaft slightly bent.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed five conforming clips.In addition the device locked out as intended.Possible causes for the damage found may be inadvertent pressure placed on the device shaft through bending, pressing against the trocar, other devices on the back table, using the device as a retractor or moving heavy tissue with the device shaft.
|