Catalog Number EL5ML
Device Problems Bent (1059); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/13/2015
Event Type  malfunction  
Event Description
It was reported that during a laparoscopic sleeve gastrectomy procedure, as per hospital documentation submitted with this document - "hand piece not working/release clips and seems like a bend above the handle." further information to be provided by sales representative when available.It is unknown what was used to complete the procedure.There were no adverse consequences for the patient reported.
 
Manufacturer Narrative
(b)(4).Should the information be provided later, a supplemental medwatch will be sent.Additional information was requested and the following was obtained: how did device not work, did clips not deploy or release? dr.Was applying clip on anastomosis site (edge) he started complaining when the instrument (he thought) went 'wobbly' or bending.Where was the bend on the device (was shaft bent)? he could not release clips when he noticed the issue.How was case completed? we had to open another one (ligamax) to proceed with the operation to achieve hemostasis.The analysis results for the el5ml device found that it was returned with the shaft slightly bent.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed five conforming clips.In addition the device locked out as intended.Possible causes for the damage found may be inadvertent pressure placed on the device shaft through bending, pressing against the trocar, other devices on the back table, using the device as a retractor or moving heavy tissue with the device shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4685201
MDR Text Key5635397
Report Number3005075853-2015-02293
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/22/2019
Device Catalogue NumberEL5ML
Device Lot NumberL4FE85
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/11/2015
Initial Date FDA Received04/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-