• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L. - CRM FACILITY ORCHESTRA; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L. - CRM FACILITY ORCHESTRA; PROGRAMMER, PACEMAKER Back to Search Results
Model Number ORCHESTRA PLUS
Device Problem Operating System Version or Upgrade Problem (2997)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/19/2015
Event Type  malfunction  
Event Description
During the smartview upgrade to 2.44 version, the message error stating about virtual memory low was displayed.The upgrade seemed to be completed.The programmer rebooted as expected at the end of upgrade process.There were no files displayed on the user interface.Preliminary analysis showed a hardware failure.The programmer was returned for repair.
 
Manufacturer Narrative
Please refer to the attached analysis report.
 
Event Description
During the smartview upgrade to 2.44 version, the message error stating about virtual memory low was displayed.The upgrade seemed to be completed.The programmer rebooted as expected at the end of upgrade process.There were no files displayed on the user interface.Preliminary analysis showed a hardware failure.The programmer was returned for repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORCHESTRA
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 9214 0
FR  92140
Manufacturer (Section G)
KONTRON SUDETENSTR. 8 87600 KAUFBEUREN GERMANY
parc d'affaires noveos 4 avenue réaumur
.
clamart 9214 0
FR   92140
Manufacturer Contact
elodie vincent
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
0146013665
MDR Report Key4689989
MDR Text Key5944210
Report Number1000165971-2015-00223
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PP980049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Invalid Data
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORCHESTRA PLUS
Device Catalogue NumberORCHESTRA PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2015
Is the Reporter a Health Professional? No
Distributor Facility Aware Date03/19/2015
Initial Date Manufacturer Received 06/11/2015
Initial Date FDA Received04/14/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/11/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-