Echo probe overheated and echo machine froze during echo study on a patient.It also has frozen when the machine was not actively being used on a patient.This was a recurring problem at the site.It occurred on 2 probes, on multiple patient uses.One of the probes exhibited the overheating issue right out of the box.Both of these 2 probes have been returned to the field service rep.The manufacturer is aware of this problem happening at other hospitals.It is possible that the overheating is due to the crystals in the probe spinning continually, which is not supposed to happen.
|