Model Number UNK509
Device Problem Physical Resistance (2578)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/09/2015
Event Type  malfunction  
Event Description
Same case as: mdr id 2134265-2015-01943.Reportable based on device analysis of related rotalink completed on (b)(4) 2015.It was reported that the catheter could not be loaded on the wire.A 1.50mm rotalink¿ plus and an unspecified rotawire were selected to treat the target lesion.However, it was not possible to insert the rotawire through the burr.The procedure was completed with another of the same device and the same rotawire.No patient complications were reported and the patient's status was good.However, device analysis revealed a fractured guidewire inside the burr.
 
Manufacturer Narrative
Device evaluated by mfr: device evaluated by mfr: the device has not been returned for analysis.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause of the reported difficulties may be due interaction with another device.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTABLATOR ROTATIONAL ATHERECTOMY SYSTEM
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4697396
MDR Text Key5714777
Report Number2134265-2015-01990
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNK509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/19/2015
Initial Date FDA Received04/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial