• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750833
Device Problems Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
A nurse reported that the tubings are loose and fall off of the handpiece during cataract procedures.According to the surgeon, this has led to collapsed anterior chambers.The procedures were completed by retightening the connection of the tubing to the handpiece.There was no report of patient harm.Additional information and a product sample have been requested.
 
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Manufacturer Narrative
This is one of two complaints reported for this finished goods lot; however this is the first complaint for this issue for this lot.Review of the device record history (dhr) indicates the order was built to specification.The customer did not retain a sample for this complaint report; visual inspection or functional testing could not be conducted.The root cause of the customer's complaint could not be established as a sample has not been received.Without a sample, it is not possible to isolate the root cause.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
eddie darton, md, jd
mail stop ab2-6
mail stop r3-48
fort worth, TX 76134
8175686660
MDR Report Key4699576
MDR Text Key5727104
Report Number2028159-2015-05670
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750833
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/05/2015
Initial Date FDA Received04/16/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/08/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ULTRASOUND, FLUIDICS MANAGEMENT SYSTEM; UNSPECIFIED HANDPIECE
Patient Outcome(s) Other;
-
-