Additional information: during inner illumination the devices' lumen was found to be blocked.After cleaning the device, it was still blocked.After slicing the device, two lumen openings, on both sides of the restriction unit were seen, unknown material was found in one of the openings.Analysis of the unknown material under sem indicated that the material is an organic substance.The organic substance probably originated from the patient's eye, as the shunt was implanted for about a year.During production, 100% final inspection is being performed on the entire batch, including inner illumination.If such a defect had been noticed during the inspection, the product would have been rejected immediately.Having said that, one may conclude, that the blockage was formed after the product had left the manufacturing plant.The root cause is external - related to patient condition / non-product factors, however, the blockage could have been caused by many different reasons during and after the clinical procedure.The blockage does not seem to be product related since lumen openings were seen on both sides of the restriction unit and no welding penetrations were found.Therefore, there is no evidence for an inherent defect that might have caused the event, and the root cause is seem to be external - related to patient condition / non-product factors.(b)(4).
|