It was reported: during the first ventilation everything was okay, but after a certain time the saturation dropped and the child became blue.According to dr.(b)(6) problems only occur in children anesthesia and completely randomly.Meanwhile, the anesthetics are already done by 2 persons, in order to avoid human error.There was no permanent damage or clinical intervention reported.
|
Device check and log analysis showed not device errors.A further analysis of user logs now ((b)(4) 2015) has shown that repeatedly changed settings have not been confirmed via the rotary knob.Therefore, these were not accepted by the device.The affected device was requested in lubeck for closer examination.Due to the analyzed log of the user's behavior, the user was offered a retaining on site by the drager sales on (b)(4) 2015.
|