The customer questioned the anti-ccp result of 57.49 u/ml for this sample, tested on (b)(6) 2015, because there was a large difference between this result and the patient's result of 11300 u/ml from "last summer".It is not known if erroneous results were reported outside of the laboratory.The sample was tested undiluted and diluted.The anti-ccp result from the undiluted sample was 56.74 u/ml.The sample was then diluted with an anti-ccp negative serum pool and tested using multiple dilution.A 2- fold dilution of the sample gave a result of 80.37 u/ml.The sample was repeated with 5-fold dilution and the result was 131.0 u/ml.The sample was repeated with 10-fold dilution and the result was >132.5 u/ml.The sample was repeated with 20-fold dilution and the result was >132.5 u/ml.The sample was repeated with 50-fold dilution and the result was 120.8 u/ml.The sample was repeated with a 100-fold dilution and the result was 72.53 u/ml.No adverse event was reported.The elecsys modular e170 serial number was not provided.Two samples from the patient were sent in for investigation.The results could not be reproduced.A specific root cause could not be identified during the investigation.A general reagent lot issue can be excluded.Product labeling documents autoantibodies are heterogeneous and this can cause non-linear dilution phenomena for certain individual samples.In addition, a sample dilution was not necessary as the result was within the measuring range.Recommended dilution is 1:2 to 1:5.The sample was diluted up to 1:100.
|