A review of complaint history, device history record, instructions for use (ifu), quality control, and a visual inspection and functional test of the returned device were conducted during the investigation.Product was returned in an opened and used condition.A visual inspection noted that the device was relatively clean and free of biological matter.The lumen and 30 5cm of shaft were intact.Adequate adhesive was found on the proximal fittings.Functional testing confirmed a leak approximately 25 2cm distal of the strain relief/shaft joint.Whatever penetrated the material was small as to not be able to detect the leak with the naked eye.This product is shipped with an instructions for use pamphlet, which states warnings, precautions and suggested instructions for use.There is no evidence to suggest that the product was not manufactured to specification, and we are inconclusive as to why this failure mode occurred.The appropriate internal personnel have been notified and monitoring for similar complaints will continue.
|