MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ULTRACISION HARMONIC ACE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ULTRACISION HARMONIC ACE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number ACE23E
Device Problems Break (1069); Loss of Power (1475); Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/14/2015
Event Type  malfunction  
Event Description
It was reported that prior to a gastrectomy procedure, the device was originally connected to the gen11 generator, after activation and putting the device into the patient, he found the generator suddenly switched off.The nurse tried to restart the generator; however, the power button was "on", while the screen was black.They tried to replace the power cord; still, the screen cannot be turn on.The nurse then connected this device to a gen04 generator, and tried to activate, however, it was working functionally for around 5 minutes.When the nurse tried to clean blade by using wet gauze, the blade was broken and dropped on the table.It is unknown what was used to complete the procedure.There were no adverse consequences for the patient reported.
 
Manufacturer Narrative
(b)(4).Information asked for but unknown or not provided during initial contact.Information is unavailable; device not returned.: no device received for analysis at time of submission of 3500a.When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.
 
Manufacturer Narrative
(b)(4).Additional information: the device was returned with the distal tip of the blade broken off and it was returned with the device.The remaining blade portion was scratched, evidence of contact with metal in or out of the operative field.The device was functionally tested with a generator.During functional testing on gen11 an alert screen was displayed and when tested on gen04 and error code was displayed.A probable cause of the device stop activating and display an alert screen or an error code is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure when used a gen11 an error code 5 or blade lockout will occur later in the procedure when using gen04.The generator system will revert to standby mode, when the audible alarm (solid tone) and visual alarm indicator appear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRACISION HARMONIC ACE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4719844
MDR Text Key5740260
Report Number3005075853-2015-02532
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K042777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/17/2016
Device Catalogue NumberACE23E
Device Lot NumberH91E7J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/19/2015
Initial Date FDA Received04/22/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR AND HANDPIECE
-
-