Catalog Number ACE23E |
Device Problems
Break (1069); Loss of Power (1475); Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/14/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to a gastrectomy procedure, the device was originally connected to the gen11 generator, after activation and putting the device into the patient, he found the generator suddenly switched off.The nurse tried to restart the generator; however, the power button was "on", while the screen was black.They tried to replace the power cord; still, the screen cannot be turn on.The nurse then connected this device to a gen04 generator, and tried to activate, however, it was working functionally for around 5 minutes.When the nurse tried to clean blade by using wet gauze, the blade was broken and dropped on the table.It is unknown what was used to complete the procedure.There were no adverse consequences for the patient reported.
|
|
Manufacturer Narrative
|
(b)(4).Information asked for but unknown or not provided during initial contact.Information is unavailable; device not returned.: no device received for analysis at time of submission of 3500a.When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.
|
|
Manufacturer Narrative
|
(b)(4).Additional information: the device was returned with the distal tip of the blade broken off and it was returned with the device.The remaining blade portion was scratched, evidence of contact with metal in or out of the operative field.The device was functionally tested with a generator.During functional testing on gen11 an alert screen was displayed and when tested on gen04 and error code was displayed.A probable cause of the device stop activating and display an alert screen or an error code is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure when used a gen11 an error code 5 or blade lockout will occur later in the procedure when using gen04.The generator system will revert to standby mode, when the audible alarm (solid tone) and visual alarm indicator appear.
|
|
Search Alerts/Recalls
|
|
|