A call to biomed was placed on a ekg cart in the emergency department with a problem description that it printed out the wrong patient information.Another cart seems have pulled the wrong patient information.The cart was taken to the main biomed shop to be checked.Multiple technical users(biomed tech, clinical engineers, risk management, etc.) were going over the procedure of selecting a patient from the order management system on the cart.A patient that did not have completed orders (the only one from the list) was selected.When the return key was pressed it picked another patient that had already been completed.A box saying that this patient has already been processed did not appear but instead a box asking if you want to proceed popped up with a different patient.In a busy environment even someone with much experience may not have caught this.After diagnostics were performed on this cart on this day we have only been able to duplicate the problem this one time.A service call was placed to ge and i was told they would be looking into this.I received a call today saying that the software engineers were able to reproduce this problem.
|