Model Number PT101
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 03/16/2015
Event Type  malfunction  
Event Description
A hospital in (b)(6) reported that the speaker of an airvo 2 humidifier was not working.This was found before use on a patient.
 
Manufacturer Narrative
(b)(4).Method: the complaint airvo humidifier was returned to fisher & paykel healthcare in (b)(4) for evaluation.The device was performance tested and the audible alarm function was checked.Results: during testing the airvo functioned normally.The visual alerts were operating, however no audible alarm was heard.The fault was traced to a faulty speaker and electrical resistance testing showed that the speaker's resistance was high or open circuit.A lot check revealed no other complaints of this nature for lot number 140714.Conclusion: the supplier of the speaker unit was notified and they have carried out an investigation.The problem has been traced to an issue with the gluing process.The supplier has taken steps to ensure that each speaker is checked following the gluing process and any found faulty are discarded.The airvo 2 user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases." and that "the unit is not intended for life support." the user manual also contains instructions on how to check the alarm system functionality and states that "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRVO2 HUMIDIFIER
Type of Device
AIRVO2 HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9494534000
MDR Report Key4725552
MDR Text Key16073956
Report Number9611451-2015-00218
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPT101
Device Catalogue NumberPT101
Device Lot Number140714005894
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2015
Initial Date FDA Received04/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-