ETHICON ENDO-SURGERY, LLC. LIGACLIP CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Device Problems Bent (1059); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/01/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Additional information : the analysis results of the er320 device found that it was received with a clip in the jaws; the clip was removed in order to inspect the jaws and they were found to be misaligned and yielded.In addition, a bag with a scissored clip was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality; during the analysis, the instrument was cycled and it fed and formed nine scissored clips due to the misaligned condition of the jaws.Possible causes for the condition of the jaws may be if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel.
 
Event Description
It was reported that during an unknown procedure there were bent and crooked staples.It was not known how the case was completed.There were no patient consequences reported.
 
Manufacturer Narrative
(b)(4).No device received for analysis at time of submission of 3500a.When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGACLIP CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4726474
MDR Text Key13363342
Report Number3005075853-2015-02596
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2019
Device Catalogue NumberER320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/15/2015
Initial Date FDA Received04/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2014
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-