Model Number MODEL 250
Device Problem Use of Incorrect Control/Treatment Settings (1126)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2010
Event Type  malfunction  
Event Description
During review of the patient's programming history, it was observed that a programming anomaly occurred.On (b)(6) 2010, a system diagnostic test was performed which changed the settings to unintended parameters indicating that the test likely did not complete.Interrogation was performed which revealed the unintended settings.However, the settings were not corrected prior to the patient leaving the clinic.
 
Manufacturer Narrative
Analysis of programming history performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4727199
MDR Text Key22062790
Report Number1644487-2015-04557
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number537101
Other Device ID NumberVERSION 7.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 03/27/2015
Initial Date FDA Received04/24/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/23/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age56 YR
-
-