The hill-rom technician found the cpr cable was disconnected from the handle.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare bed components are functioning as originally designed.Examine the handles, cables, and cpr mechanism on the head motor.Make sure the screws are installed and fully tightened.Examine the condition of the two cpr release handles, cpr cable, and cpr mechanism on the head motor.Operate the head section to the high position, then engage one of the cpr release.Make sure the head section lowers.Adjust the cpr cable as necessary.Do the same tests on the other side of the bed.Release the cpr handles and make sure the mechanism locks correctly when you operate the head up control for a few seconds.The head section must rise.Replace the head section motor in the event of a malfunction of the cpr cable if broken.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2015.It is unk if the facility performed any other preventative maintenance on this bed.The technician reconnected the cable to resolve the issue.Based on this info, no further action is required.
|