• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 3200
Device Problems Sticking (1597); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/27/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from a hill-rom technician stating the head section would not lower when cpr lever was used.The bed was located in the bed shop at the account.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hill-rom technician found the cpr cable was disconnected from the handle.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare bed components are functioning as originally designed.Examine the handles, cables, and cpr mechanism on the head motor.Make sure the screws are installed and fully tightened.Examine the condition of the two cpr release handles, cpr cable, and cpr mechanism on the head motor.Operate the head section to the high position, then engage one of the cpr release.Make sure the head section lowers.Adjust the cpr cable as necessary.Do the same tests on the other side of the bed.Release the cpr handles and make sure the mechanism locks correctly when you operate the head up control for a few seconds.The head section must rise.Replace the head section motor in the event of a malfunction of the cpr cable if broken.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2015.It is unk if the facility performed any other preventative maintenance on this bed.The technician reconnected the cable to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4731080
MDR Text Key5581177
Report Number1824206-2015-00529
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/27/2015
Initial Date FDA Received04/22/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-