Discordant, (b)(6) results were obtained on two patient sample on an advia centaur xp instrument.The discordant results were not reported to the physician(s).The sample were repeated on an alternate advia centaur xp instrument.The corrected results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, (b)(6) results.
|
A siemens customer service engineer (cse) was dispatched to the customer site.The cse changed the tubing for the sample probe plunger, lubricated the syringe, removed and cleaned the plunger, changed the vertical probe belt, changed the acid and base pumps, and recalibrated the sample probe bottom.The customer is running siemens (b)(6) quality controls every four hours in replicates of ten.The cause of the discordant, (b)(6) result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|