Olympus was informed that after the completion of a diagnostic cystoscopy procedure, it was observed that tissue fibers were coming out of the tip of the scope at the distal end.It was also reported that some of the distal end cover glue had fallen into the patient and was not retrieved by the physician.The physician believed that the device fragments will be flushed out naturally.There was no patient injury reported and intended procedure was completed using same device.It was further reported that during previous cleaning of the device there were several pieces of the black adhesive at the distal end of the device missing.No further information has been provided.
|