It was reported that the patient's generator was replaced due to end of service on (b)(4() 2015.During the surgery, it was mentioned that the surgeon had difficulty removing the lead from the to be explanted generator.As a result, the header on the explanted generator broke during this explant process.It was confirmed that the header was not broken prior to the surgery.Diagnostics performed with the old generator and newly implanted generator were reported to be within normal limits.The explanted generator was returned to the manufacturer.Analysis is underway but has not been completed.
|
Based on the location of the tool marks observed on the pulse generator case and header, the header detachment was determined to have most likely occurred during or after the explant process.Other than the header observation, there were no performance or any other type of adverse conditions found with the pulse generator.
|