Model Number 106A3
Device Problems Connection Problem (2900); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/20/2015
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the user observed a coaxial connection icon on the cryoconsole screen.This occurred in preparation for a cryoablation procedure (prior to system flush, after the integrity test).He issue resolved after tapping on check valve cv4 and the cryoablation procedure was completed.No patient complications were reported.Technical support was contacted and a field service visit was recommended.Reportable following revision to regulatory reporting criteria.
 
Manufacturer Narrative
Technical support was notified of this occurrence.Upon review of complaint information, a service order was initiated to review the console performance on site.The reported issue was not reproduced by field service engineering; but the check valve (cv4) was replaced as a precautionary measure due to the specific nature of the issue reported during the case.The flow meter was also adjusted during the service visit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOCONSOLE
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5H3
CA   H9H 5H3
Manufacturer Contact
voula radiotis
16771 chemin ste-marie
kirkland,qc H9H 5-H3
CA   H9H 5H3
5146941212
MDR Report Key4753487
MDR Text Key5824658
Report Number3002648230-2015-00135
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number106A3
Device Catalogue Number106A3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2015
Initial Date FDA Received05/06/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-