|
Model Number H749518080 |
Device Problems
Fluid/Blood Leak (1250); Flushing Problem (1252); Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/26/2015 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on (b)(4) 2015.It was reported that imaging catheter leak, kink and inability to flush occurred.During procedure, an opticross¿ imaging catheter was selected to view the severely tortuous target lesion which was obtained via femoral artery.After the opticross¿ was delivered a few times, it was then flushed outside the patient.However, it was noted that the saline leaked from a part of the shaft.It was further noted that the device was kinked.Subsequently, the device became unable to flush.The procedure was completed with another of the same device.No patient complications were reported.The patient's status was good.However, device analysis revealed a lap joint partial separation.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the complaint device was returned for evaluation.Evaluation of the returned device revealed a kink in the sheath (lap joint section) at 73cm from the femoral marker to the distal end and a lap joint partial separation in the device.The telescope assembly was not able to properly pull back, advance, or retract due to the lap joint partial separation.During flushing, fluid was leaking at the sheath lap joint junction between the blue sheath and clear imaging window tubing.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
|
|
Search Alerts/Recalls
|
|
|