• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; APS 1 (HEART L8UNG CONSOLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; APS 1 (HEART L8UNG CONSOLE) Back to Search Results
Model Number 801764
Device Problem Failure to Run on Battery (1466)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/15/2015
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the perfusion system would not switch over to battery power.There was a "battery needs service" error message observed.The device was not changed out, since they have a universal power supply (ups) in the operating room, they went on with the case.This was an off pump procedure.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the pt.
 
Manufacturer Narrative
Eval is in progress, but not yet concluded.The device was charged overnight.The total nominal battery capacity shows 18 amp hours (ah), power mode-2, power manager 1-0 (good) and power manager 2-0 (good).Per the distributor, when the system was unplugged there was 160 minutes displayed on the central control lights on the ccm and front panel of the system were both green.The distributor checked the system and found the power manager board was defective.They replaced with a new board from company stock.The system is now working fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
APS 1 (HEART L8UNG CONSOLE)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4762821
MDR Text Key5786097
Report Number1828100-2015-00396
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801764
Device Catalogue Number801764
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/30/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/21/2015
Initial Date FDA Received05/07/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-