MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FREE PSA; TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FREE PSA; TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA Back to Search Results
Catalog Number 03289788190
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/30/2015
Event Type  malfunction  
Event Description
The customer received questionable total psa total (free + complexed) psa - prostate-specific antigen (total psa) and free psa results for one patient sample from cobas e601 analyzer serial number (b)(4).The specific date of testing was not provided.The initial free-psa result was 1.64 ng/ml and the initial total psa result was 0.902 ng/ml.The customer recentrifuged the sample and repeated testing with the same results.(b)(4).Information concerning if any erroneous result was reported outside the laboratory was requested, but was not provided.No adverse event was reported.
 
Manufacturer Narrative
This event occurred in (b)(6).
 
Manufacturer Narrative
A specific root cause could not be identified as no patient sample could be provided for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREE PSA
Type of Device
TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 6830 5
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key4768797
MDR Text Key5778201
Report Number1823260-2015-03455
Device Sequence Number1
Product Code MTG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P000027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03289788190
Device Lot Number178872
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received 04/30/2015
Initial Date FDA Received05/13/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age065 YR
-
-