• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Unable to Obtain Readings (1516); Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/29/2015
Event Type  malfunction  
Event Description
It was reported that the impedance values were varying intra-operatively on the day of report.After multiple trouble shooting attempts, the lead was wiped and impedances tested at higher values, the patient got motor response as well as in range values for all the bipolar pairs except 0 <(>&<)>1.The case values all showed ¿???¿.Additional information reported the issue appeared to be resolved.No further information was provided.There was no patient harm reported.
 
Manufacturer Narrative
Concomitant medical products: product id: neu_unknown_lead, product type: lead.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 120
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4773259
MDR Text Key5952473
Report Number3007566237-2015-01321
Device Sequence Number1
Product Code EZW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/29/2015
Initial Date FDA Received05/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-