• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON UP-DRAFT II OPTI-NEB NEBULIZER W/RESERV; SMALLL VOLUME NEBULIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON UP-DRAFT II OPTI-NEB NEBULIZER W/RESERV; SMALLL VOLUME NEBULIZER Back to Search Results
Catalog Number 1734
Device Problems Detachment Of Device Component (1104); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/27/2015
Event Type  malfunction  
Event Description
The customer alleges that the tubing is popping off at both ends during treatment.No report of a patient injury or consequence.
 
Manufacturer Narrative
Qn#(b)(4).It is unknown if the device sample is available for evaluation.A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product or a picture of the defect was not provided.A device history record review could not be conducted since the lot number was not provided.Customer complaint cannot be confirmed since the sample and lot number was not provided for investigation.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility nor is being manufactured at the time.If the sample becomes available this investigation will be updated with the evaluation results.
 
Manufacturer Narrative
(b)(4).The lot number was originally reported as unknown; however, the lot number of the sample received was 74a1503034.A device history record (dhr) review was performed and there were no issues found that could relate to the reported complaint.The sample was returned for evaluation.A visual exam and functional testing was performed and no issues were found.No "popping off" issues were observed during nebulization as reported in the complaint.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
 
Event Description
The customer alleges that the tubing is popping off at both ends during treatment.No report of a patient injury or consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON UP-DRAFT II OPTI-NEB NEBULIZER W/RESERV
Type of Device
SMALLL VOLUME NEBULIZER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
parque industrial finsa
nuevo laredo 8827 5
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key4774464
MDR Text Key19408147
Report Number3004365956-2015-00148
Device Sequence Number1
Product Code CCQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number1734
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/24/2015
Initial Date FDA Received05/14/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/29/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-